Key Features
- Adopts 1200 bar/mm high-performance grating. New light control system makes the instrument light switch more quickly.
- The instrument adopts 320*240 lattice high-lighted LCD display, direct display scan map, the screen interface simple, convenient use.
- Socket xenon lamp and tungsten lamp design, after change light no need optical debugging.
Large sample chamber, can accommodate 5-100mm various specifications of the cuvette. - Can be directly connected to the printer to print patterns and the experimental data.
Optional disk automatic 8 joint cuvettes holder.
Technical Specifications
| Optical system: | Czerny-turner CT grating monochromatic, 1200/mm |
| Wavelength range: | 320-1100 nm |
| Wavelength accuracy: | 0.3 nm |
| Wavelength repeatability: | <=0.2nm |
| Measurement range: | -03A-3A |
| Photometric accuracy: | <= 0.5%T |
| Photometric repeatability: | <= 0.3%T |
| Spectral bandwidth: | 2 nm |
| Zeroing: | Automatic |
| Stray light: | <= 0.05%T |
| Stability: | +/- 0.001 A/h @500 nm) |
| Display modes: | 320*240 lattice high-lighted LCD display |
| Scanning speed: | Fast, Mid, Slow |
| Data output (print): | USB Export |
| Light source: | Tungsten Lamp (6V10W) |
| Input voltage: | AC 220V/50Hz or AC110V/60Hz |