- Measure meteorological parameters:wind speed, wind direction, rainfall, radiation, temperature, barometric pressure, relative humidity
- Measures air pollutants: SO2, NOx, CO, O3, TVOC, H2S, NH3, CO2, HCHO etc
- Measure particles: PM1, PM2.5, PM10, TSP
- Modular design, arbitrary configuration combination
- On-site calibration according to traceable reference standards
- Unique light scattering particle sensors. Dynamic heating and dehumidification control was adopted for particle sampling to remove the influence of water mist on the measured data.
- High-precision ADC data acquisition system, capable of monitoring PPB level pollutants;
- Industrial-grade embedded processor, suitable for harsh outdoor environment
- Automatically upload real-time data to cloud platform by 4G. Data can be published on the Internet for easy public access and to raise awareness of current air pollution levels with cloud platform
- Free RS485 Modbus protocol is available
- Removable 32G SD Card (.csv file)
- Solar power supply is available. Equipped with solar panel controller
- Outdoor design with housing to prevent dust and rain, IP53
- Touch screen , easily read data and make operation
- can set the data storage time interval : 1min, 5min, 10min, 30min, 60min.
- can download historical data by minutes/hours/days in cloud platform
- Alarm when detection values trigger the threshold
Technical Specifications:
Particulate matter parameters
| Sensor | Light scattering sensor |
| Diameter | PM2.5, PM10 |
| Measured
Range |
PM2.5 0-1000ug/m3
PM10 0-1000ug/m3 |
| Resolution | 1ug/m3 |
| Sampling method | Pumped |
| Heating | Automatic |
| Accuracy | >100ug/m3, ±25% |
| Calibration | Automatic |
Gas parameters
| Gas | Sensor type | Range | Resolution | Accuracy | T90 | Min Test Value |
| O3 | Electrochemical | 500 ppb | 1 ppb | ±10%F.S | ≤120 s | 5ppb |
| CO | Electrochemical | 50 ppm | 0.1 ppm | ±10%F.S | ≤90 s | 0.04ppm |
| SO2 | Electrochemical | 500 ppb | 1 ppb | ±10%F.S | ≤90 s | 5ppb |
| NO2 | Electrochemical | 500 ppb | 1 ppb | ±10%F.S | ≤90 s | 5ppb |
| CH4 | IR | 5%VOL | 0.01%VOL | ±10%F.S | ≤90 s | 0.01%VOL |
| VOC | PID | 10ppm | 0.1ppm | ±10%F.S | ≤90 s | 0.04ppm |
Meteorological parameters
| Parameter | Range | Resolution | Accuracy |
| Temperature | -22°F to 122°F (-30 ~ 50 ℃) | 0.18°F (0.1℃) | 0.18°F (±1 ℃) |
| Relative humidity | 0~95% RH | 1% RH | ±5% RH |
| Barometric pressure | 600 ~ 1100 hpa | 1hPa | ±0.5 hPa |
Other parameters
| Parameter | Specification |
| Power Requirements | Input Voltage: 12VDC |
| Power Consumption: Max 80W | |
| Power Supply | 220V, 50Hz (Solar power optional) |
| Preheat Time | 90 seconds |
| Protection Level | IP53 |
| Humidity | 10–95% RH (non-condensing) |
| Operating Environment | Temperature: -4°F to 122°F ( -20°C to 50°C) |
| Dimensions | Standard: 17.32 in × 7.87 in × 26.38 in (440mm × 200 mm× 670 mm )(L × W × H)
750×200×1070mm(L×W×H) (Max, with wind and PM detection) |
| Weight | 55.12 lb (25 kg) |
| Warranty time | 12 months |






Reviews